# Equipment Setup
|
|
We got our pulse height to be 3 V. 2.4 was not super attainable. Oh well. PZ was adjusted
|
|
|
|
Once at 20microsecond timescale we noticed some undershoot going on. Pole zero adjustment was modified to compensate for this.
|
|
|
|
We got a spectrum that we recorded for 1 minute. It looks like 5.9 after changing vertical scale to "A" for auto.
|
|
|
|
We had to adjust 11to 595.6V on the high voltage power supply to get our spectrum in the right place.
|
|
|
|
We set the lower level discretization to 130. 75 still incorporated background noise.
|
|
|
|
Accurate coefficients are in the cobalt ASCII file.
|
|
|
|
First rip on the unknown sample, 616 counts at 658kev
|
|
|
|
Exercise 5.7
|
|
Best guess rubidium at 657.777 134 cA gross area 2507
|
|
With Cs source now, gross area 6886
|
|
background count gross area 81
|
|
|
|
Range Table
|
|
Hubert Binne
|