# Equipment Setup
|
||||||
|
We got our pulse height to be 3 V. 2.4 was not super attainable. Oh well. PZ was adjusted
|
||||||
|
|
||||||
|
Once at 20microsecond timescale we noticed some undershoot going on. Pole zero adjustment was modified to compensate for this.
|
||||||
|
|
||||||
|
We got a spectrum that we recorded for 1 minute. It looks like 5.9 after changing vertical scale to "A" for auto.
|
||||||
|
|
||||||
|
We had to adjust 11to 595.6V on the high voltage power supply to get our spectrum in the right place.
|
||||||
|
|
||||||
|
We set the lower level discretization to 130. 75 still incorporated background noise.
|
||||||